Pulse oximeters utilized in steady sufferers hospitalized in an intensive care unit (ICU) had reasonable accuracy when in comparison with arterial blood gasoline measurements, in keeping with research findings revealed in Respiratory Care.
The Oxygap research evaluated the accuracy and biases of 4 pulse oximeters which can be ceaselessly utilized in medical follow by evaluating the SpO2 values generated by these devices with SaO2 values measured through drawing arterial blood gases.
Members had been grownup sufferers admitted to an ICU in Quebec Metropolis, Canada, from March 24 to September 17, 2021. All sufferers had been steady relating to oxygen administration parameters (ie, no change in fraction of impressed oxygen [FiO2] and oxygen circulate for the final half-hour) and hospitalized with an arterial catheter in place.
The oximeters that had been assessed had been: (1) Nonin within the FreeO2 system (Oxynov); (2) Masimo (nontouch display screen, Radical-7); (3) Philips (FAST); and (4) Nellcor (N-600). SpO2 values for every pulse oximeter had been obtained concomitantly to the arterial blood gasoline. The first outcomes had been SpO2 values for every pulse oximeter and SaO2 values from the measurement of arterial blood gases.
A complete of 211 units of concomitant oxygen saturation measurements (with every set together with 1 SaO2 worth and 4 SpO2 values, 1 for every of the 4 pulse oximeters) had been obtained from 193 contributors. Of these units of measurements, 153 (73%) had been from males and 58 (27%) had been from ladies. The sufferers had a imply (SD) age of 66 (11) years, and most had “gentle pores and skin,” famous the researchers (ie, 96.2% had Fitzpatrick scale sort 1 or 2 pores and skin pigmentation).
“
Oximeters needs to be used realizing their limits, and when SpO2 is utilized by clinicians to focus on the oxygen circulate or FiO2, the oxygenation targets ought to most likely be tailored primarily based on the oximeter and pores and skin pigmentation.
The contributors had a imply SaO2 worth of 93.6% (2.2%). The Philips oximeter had a constructive bias of 0.9%. Destructive biases had been noticed for the Nonin, Nellcor, and Masimo oximeters (-3.1%, -0.3%, and -0.2%, respectively).
In contrast with the concomitant SaO2 measurements, SpO2 was underestimated in 55% of instances. Among the many 4 pulse oximeters evaluated, an total frequency of constructive and adverse variations of 4% or extra between SaO2 vs SpO2 measurements had been present in 35.2%, 2.4%, 8.6%, and 9.5% of the Nonin, Nellcor, Masimo, and Philips pulse oximeter measurements, respectively.
Relating to accuracy, the heart beat oximeters confirmed massive random errors. The Nellcor, which had an R2 coefficient worth of 0.5148, was discovered to have barely larger accuracy than the opposite pulse oximeters, which had R2 coefficients of 0.4306 (Nonin), 0.4328 (Masimo), and 0.4387 (Philips).
The heart beat oximeters’ potential to establish hypoxemia assorted. Amongst cohort contributors, 19 sufferers had a PaO2 worth from 55 mmHg to 60 mmHg, and 12 had SaO2 values from 86% to 90%, indicating reasonable hypoxemia. The Nonin pulse oximeter was essentially the most delicate for detecting hypoxemia, detecting 100% of the states of reasonable hypoxemia, whereas the Philips pulse oximeter detected 11%, the Masimo detected 37%, and the Nellcor detected 26% of hypoxemic occasions.
Research limitations embrace the low variety of pores and skin pigmentation in research contributors and the shortage of accounting for bias launched by darkish pores and skin pigmentation in among the outcomes; small pattern measurement; and a restricted variety of sufferers with SaO2 values of lower than 90%.
“Oximeters needs to be used realizing their limits, and when SpO2 is utilized by clinicians to focus on the oxygen circulate or FiO2, the oxygenation targets ought to most likely be tailored primarily based on the oximeter and pores and skin pigmentation,” acknowledged the researchers.
Disclosure: A few of the research authors declared affiliations with a medical tools producer. Please see the unique reference for a full checklist of authors’ disclosures.